Pull the implant out using a proper screwdriver, either a manual screwdriver or a ISO screwdriver.
Put into place and partly screw down, continue screwing with the handpiece or with the ratchet P500002, using the appropriate ISO insert P900000 and the appropriate screwdrivers P500008, long, four-lobed cam, or P500011 short, four-lobed cam.
To screw the implant fixture, it is possible to use the ratchet with big insert P800000, using the manual screwdrivers for three-lobed camP 600008 long, and P600011 short.
The cam screwdrivers have dot markings used as references for directing the cams of the implant while screwing. This should be useful when applying the prosthesis, if we are using an angled abutment.
Recommended screwing speed: 15/20 rpm.
Torque not exceeding 40 Ncm.
No irrigation.
We recommend removing the cap screw from its housing and screwing it onto the implant using the manual screwdriver with no torque control P500005/ P600000, to have direct perception of correct positioning of the screw and of its screwing.
Pull the implant out using a proper screwdriver, either a manual screwdriver or a ISO screwdriver.
Put into place and partly screw down, continue screwing with the handpiece or with the ratchet P500002, using the appropriate ISO insert P900000 and the appropriate screwdrivers P500008, long, four-lobed cam, or P500011 short, four-lobed cam.
To screw the implant fixture, it is possible to use the ratchet with big insert P800000, using the manual screwdrivers for three-lobed camP 600008 long, and P600011 short.
The cam screwdrivers have dot markings used as references for directing the cams of the implant while screwing. This should be useful when applying the prosthesis, if we are using an angled abutment.
Recommended screwing speed: 15/20 rpm.
Torque not exceeding 40 Ncm.
No irrigation.
We recommend removing the cap screw from its housing and screwing it onto the implant using the manual screwdriver with no torque control P500005/ P600000, to have direct perception of correct positioning of the screw and of its screwing.
CODE
PRODUCT NAME
DIAMETER MM
LENGTH MM
COLOR CODE
IVA
P125011
FIXTURE PLATFORM SWITCHING
5 external body
11
green internal connection 4.5mm diameter
4%
Sign up for Email