PRIMER SR FOUR-LOBED CAM, LARGE Ø5

PRIMER SR FOUR-LOBED CAM, LARGE Ø5 | fixture dental internal external connection immeditate loading implant italy

PRIMER SR FOUR-LOBED CAM, LARGE Ø5
Ask
Print
Invite a Friend

PRIMER SR FOUR-LOBED CAM, LARGE Ø5

Pull the implant out using a proper screwdriver, either a manual screwdriver or a ISO screwdriver.
  • Pull the implant out using a proper screwdriver, either a manual screwdriver or a ISO screwdriver.
  • Without touching the fixture, put into place and partly screw down. Continue screwing with the handpiece or with the ratchet P500002 using the appropriate ISO insert P900000 and the appropriate screwdrivers:  P500009, long, four-lobed cam, or P500012 short, four-lobed cam.
  • To screw the implant fixture, it is possible to use the ratchet with big insert P800000, using the manual screwdrivers for four-lobed cam P600012 short and P600009 long.
  • The cam screwdrivers have dot markings used as references for directing the cams of the implant  whilescrewing. This should be useful in the prosthetic stage, if we are using an angled abutment.
  • Recommended screwing speed: 15/20 rpm.
  • Torque not exceeding 40 Ncm.
  • No irrigation.
  • We recommend removing the cap screw from its housing and screwing it onto the implant using the manual screwdriver with no torque control,  P500005 / P600000  to feel directly the correct positioning of the screw and how it is screwed in.

 

  • Pull the implant out using a proper screwdriver, either a manual screwdriver or a ISO screwdriver.
  • Without touching the fixture, put into place and partly screw down. Continue screwing with the handpiece or with the ratchet P500002 using the appropriate ISO insert P900000 and the appropriate screwdrivers:  P500009, long, four-lobed cam, or P500012 short, four-lobed cam.
  • To screw the implant fixture, it is possible to use the ratchet with big insert P800000, using the manual screwdrivers for four-lobed cam P600012 short and P600009 long.
  • The cam screwdrivers have dot markings used as references for directing the cams of the implant  whilescrewing. This should be useful in the prosthetic stage, if we are using an angled abutment.
  • Recommended screwing speed: 15/20 rpm.
  • Torque not exceeding 40 Ncm.
  • No irrigation.
  • We recommend removing the cap screw from its housing and screwing it onto the implant using the manual screwdriver with no torque control,  P500005 / P600000  to feel directly the correct positioning of the screw and how it is screwed in.

 

Content of the pack:
- Threaded implant Grade 4 Ti
- Mounter Grade 5 Ti
- Cap screw Grade 2 Ti
CODEPRODUCT NAMEDIAMETER MMLENGHTCOLOR CODEIVA
P105085FIXTURE58.5green4%
P105010FIXTURE59.95green4%
P105011FIXTURE511green4%
P105013FIXTURE513green4%
P105015FIXTURE515green4%