All E.T. dental Implants have a mounter for manual insertion. The implant is extracted from the sterile vial using the mounter provided and inserted in position and partially screwed; then continue screwing by using a Ratchet (code number P500002) eq
  • All E.T. dental Implants have a mounter for manual introduction.
  • Once the mounter has been grasped between the fingers, without touching the fixture, put into place and partly screw down, continue screwing  using the handpiece or the Ratchet P500002 equipped with the large Insert P800000 with the manual screwdrivers: Long manual Screw-driver P600008 for small four-lobed cam implant and Short manual Screw-driver P600011 for small four-lobed cam implant.
  •  Screw-drivers have dot markings serving used as references for directing the cams of the implant while screwing. This should be useful when applying the prosthesis, if we are using an angled abutment.
  • Recommended speed when using the handpiece: 15 to 20 rpm
  • Maximum torque: 40 N. 
  • Remove the cap screw from the mounter head and screw it onto the implant using the long manual hexagonal screwdriver P600006 or short manual hexagonal screwdriver P600007 in order to have a correct positioning of the implant.

 

  • All E.T. dental Implants have a mounter for manual introduction.
  • Once the mounter has been grasped between the fingers, without touching the fixture, put into place and partly screw down, continue screwing  using the handpiece or the Ratchet P500002 equipped with the large Insert P800000 with the manual screwdrivers: Long manual Screw-driver P600008 for small four-lobed cam implant and Short manual Screw-driver P600011 for small four-lobed cam implant.
  •  Screw-drivers have dot markings serving used as references for directing the cams of the implant while screwing. This should be useful when applying the prosthesis, if we are using an angled abutment.
  • Recommended speed when using the handpiece: 15 to 20 rpm
  • Maximum torque: 40 N. 
  • Remove the cap screw from the mounter head and screw it onto the implant using the long manual hexagonal screwdriver P600006 or short manual hexagonal screwdriver P600007 in order to have a correct positioning of the implant.

 

Content of the pack:
- Threaded implant Grade 4 Ti
- Mounter Grade 5 Ti
CODEPRODUCT NAMEDIAMETER IN MMLENGHTCOLOUR CODEIVA
P114213EASY TAPPING FIXTUREØ 4.213 mmblue4
P114215EASY TAPPING FIXTUREØ 4.215 mmblue4
P114218EASY TAPPING FIXTUREØ 4.218 mmblue4
P114211EASY TAPPING FIXTUREØ 4.211 mmblue4%